When the biochemical incubator is in use, if there is a "no cooling" fault phenomenon, self-exclusion and treatment methods
If the biochemical incubator is not cooled, self-removal and treatment methods:
If the biochemical incubator is not cooled, there are four general faults:
1. The position of the transfer switch is incorrect;
2. The compressor is broken or the pipeline is blocked or R12 is leaking;
3. The compressor overheat protector starts;
4. The compressor protector is burnt out (with a burnt smell).
There are several corresponding methods to exclude and solve the above 4 points:
1. The position of the transfer switch is not correct; treatment method: set the correct position
2. The compressor is broken or the pipeline is blocked or R12 is leaking; treatment methods: change, remove the blockage, and add R12
3. The compressor overheating protector starts; treatment method: wait for the compressor temperature to drop, and recover naturally
4. The compressor protector is burnt out (with a burnt smell). Treatment method: replace the protector.