When the biochemical incubator is in use, if there is a "no cooling" fault phenomenon, self-exclusion and treatment methods

Article source:LAWSON Author:Lawson Popularity: 222 Issuing Time:2018/3/13 9:15:23

If the biochemical incubator is not cooled, self-removal and treatment methods:

If the biochemical incubator is not cooled, there are four general faults:

1. The position of the transfer switch is incorrect;

2. The compressor is broken or the pipeline is blocked or R12 is leaking;

3. The compressor overheat protector starts;

4. The compressor protector is burnt out (with a burnt smell).

There are several corresponding methods to exclude and solve the above 4 points:

1. The position of the transfer switch is not correct; treatment method: set the correct position

2. The compressor is broken or the pipeline is blocked or R12 is leaking; treatment methods: change, remove the blockage, and add R12

3. The compressor overheating protector starts; treatment method: wait for the compressor temperature to drop, and recover naturally

4. The compressor protector is burnt out (with a burnt smell). Treatment method: replace the protector.