Proper use steps of sterile homogenizer

Article source:LAWSON Author:Lawson Popularity: 346 Issuing Time:2018/6/13 9:08:19

Aseptic homogenizer pedal distance adjustment function:

Thick/thin samples can be completely crushed

Prompt function: (after the crushing is completed, a beep will be emitted automatically)

The function of checking the operation status, speed, and time by LCD

Continuous reading & 1 second to 60 minutes, freely adjustable

Implement minutes:seconds or hours:minutes unit selection (can be changed internally)

When the run is complete, beep (adjustable time)

When the power failure recovers, it will run according to the remaining time (time memory function: stored every 2 minutes)

When homogenization starts, the green light flashes

When the homogenization is over, the red light is on

Aseptic homogenizer box door: strengthen the transparent window, can confirm the crushing operation degree The stamping function of the pedal makes the crushing more detailed

The correct use of the sterile homogenizer is as follows:

1. Put the sample to be processed into a sterile homogeneous bag.

2. Open the door of the homogenizer (it can be fully opened), and clamp the opening of the sterile bag when closing the door of the homogenizer.

3. Set the application program (such as slap speed, slap time, etc.) to perform slap homogenization.

4. After the experiment is completed, open the door of the homogenizer and take out the sample.

The working principle of the sterile homogenizer is:

Put the original sample (the large one needs to be cut into blocks of about 10×10mm), put it into the homogenizing bag with some liquid or solvent, and repeatedly hammer the sample homogenizing bag by the hammering plate of the instrument to generate pressure, cause Shake and accelerate the mixing so that the microbial components in the solution are uniformly distributed. This instrument is suitable for the homogenization of brain, kidney, liver, spleen and other tissues and microorganisms, and is widely used in the fields of food, medicine, clinical, molecular, toxin and bacteria detection. The device is equipped with a fully open glass transparent window door, which is easy to clean, and the glass transparent window is easy to observe. The pedal distance adjustment function of the aseptic homogenizer adopts a large-screen LCD display, and the working time and homogenization speed can be intelligently adjusted. It can effectively separate the surface of the experimental sample and the homogeneous sample of microorganisms contained in it. The sample is isolated and operated by a disposable sterile homogeneous filter bag to ensure hygiene and safety, no contact with the instrument, no sample leakage and no need for system cleaning. It has the characteristics of convenience and rapidity, accurate results, soft homogenization, no pollution, no damage, no heating, no need for sterilization, no need to wash utensils, and good repeatability. This product is also suitable for the homogenization of tumor tissues (such as colorectal cancer, gastric cancer, liver cancer), and a large number of single cells can be obtained. If necessary, the tissue cells can be softly disrupted by extending the homogenization time. Treating these problems correctly can make the homogenizer serve you better.

1) The power plug must be firmly inserted in place. If it is loose, it may affect the computer controller and cause a crash. If a crash occurs, please turn off the power switch on the rear side, and restart it after 3 minutes of shutdown.

2) Please do not open the door of the homogenizer at will when the hammering plate is working to avoid overflow of the sample liquid. Should be set according to "on/off", the equipment will automatically stop running.

When the door is closed, press "Open/Stop" again, and the device will automatically complete the rest of the work. Do not reverse the operation.

3) If the instrument is not used for a long time, the power should be cut off and the plug should be removed. Prevent the aging of the electronic components inside the sterile homogenizer.

4) The bottom of the sterile homogenizer pedal distance adjustment function door is designed to prevent accidental rupture of the homogenizer bag and facilitate cleaning of spills. The bottom is designed to be empty, and a water collecting tray can be placed, so do not put your hands into the bottom when the instrument is running. , so as not to injure your fingers.

When the homogenizer is working, it is best not to open the door of the homogenizer, causing unnecessary losses.

5) Before the machine is running, please carefully check whether there are foreign objects in the homogenizing box, so as to avoid malfunction and damage to the homogenizing bag during operation.

6) Hard and sharp substances such as lumps, bones, and ice should not be used, so as not to damage the homogeneous bag.

7) The storage of the instrument and the homogenizing bag should avoid direct sunlight, especially the homogenizing bag should be stored in a place without sunlight or ultraviolet rays to avoid aging.

8) When the amount is small, when the speed needs to be accelerated, when the homogeneous fiber is relatively tough, the back knob can be used to adjust the distance between the slap plate and the viewing window to achieve a better homogeneous effect.

Pay attention to the adjusted distance to avoid damage to the sterile homogenizer such as air strikes.