LAWSON dry goods! On the Importance of Maintenance and Safety of Grinding Instruments

Article source:LAWSON Author:Lawson Popularity: 752 Issuing Time:2018/10/23 9:07:32

As the multi-sample tissue grinder slowly replaces the traditional manual grinder, the number of times the grinding machine is used in the experiment increases accordingly. In the process of using the multi-sample tissue grinder Hou should pay attention to what safety factors.

Let's talk about LAWSON LAWSONDHFSTPRP series multi-sample tissue grinderWorking principle and operation method.



  I. The working principle of LAWSON Luochanduo sample tissue grinder  

The grinding beads in the centrifuge tube are driven by mechanical transmission Make irregular movements left and right or up and down, so that the grinding beads keep hitting the sample, so as to achieve the ideal effect of crushing.



  II. Operation process of LAWSON grinding machine  

1. Select the appropriate adapter according to the sample size.

2. Put the sample into the EP tube.

3. Add grinding beads.

4. If the samples are difficult to handle, they can be placed in liquid nitrogen for freezing.

5. Place the EP tube in the adapter.

6. Install the adapter into the spindle and fasten it with the locking nut.

7. Set the grinder parameters and start.



  Three. Problems during the operation of LAWSON Grinder  

1. Adapter, the probability of this problem is relatively small.

2. Put the sample into the EP tube.

This is the first issue we should pay attention to: LAWSON EP pipe. The EP pipe must have a thick wall so that it can withstand the impact. Several empty pipe tests can be done before confirming the use of the pipe.

3. Grinding beads, this only needs to be selected according to the sample. It is not easy to be too large, and it is easy to break the tube; it is not easy to be too small, and the force is not enough, resulting in insufficient grinding.

4. Liquid nitrogen freezing treatment, pay attention to protection to avoid liquid nitrogen frostbite.

5. Place the EP tube in the adapter. When placing too much, pay attention to the balance. During the high-frequency oscillation process, if the grinder is not flat to the left and right, and the center column moves up and down, there will be offsets, which will affect the service life of the instrument, and there are potential safety hazards in the process of correct operation and use.

6. Install the adapter into the main shaft, fasten the lock and tighten the nut. Be sure to tighten the fastening nut. If it is not tightened, the adapter will impact up and down, which may easily cause a column column accident.

7. Set the parameters of the grinder. As long as you pay attention to this, the time should not be too long, such as half an hour, 1 hour.

As long as we take care to avoid these points of error, we can protect the grinder and protect our samples. If you have any doubts during use, please contact our after-sales staff, please do not disassemble and maintain by yourself!



  Fourth, the common fault analysis and troubleshooting method of LAWSON Luoshan grinding machine  

1. Click to start cannot start normally.

Please check the sensor switch at the opening to see if it is pressed. Whether the set frequency is too low.

2. After startup, there is a loud crash sound in the device.

Please check that the bottom shipping retaining nut is removed. Check if the adapter lock is vertical.

3. After the power is turned on, the program displayed on the screen will display the words "communicating"

Please check whether the data cable on the back of the display is loose.

4. During use, the adapter and the gland have unstable shaking phenomenon.

Indicates that the screw cap of the fixed adapter may be worn out, so replace it with a new screw cap.

5. The power switch light is on, but the display does not display.

It means that the display connector of the device is loose, please contact the after-sales personnel for maintenance.

6. During use, or after not using for a long time, after confirming that the power supply is correctly connected and the startup operation is correct, the device does not respond.

It means that the device has activated the self-protection function in the case of a short circuit, and you need to contact the after-sales personnel to replace the fuse.

7. The main shaft of the fixed adapter in the instrument is obviously loose or broken, or the internal parts of the instrument are faulty. Please contact our after-sales staff.


www.tissuy.com


--------------------------------------------------------- -------------